ANALISIS HASIL PENGUJIAN DAN KALIBRASI PARAMETER BLOOD PRESSURE PADA ALAT KESEHATAN BLOOD PRESSURE MONITOR

Authors

  • Nur Hadziqoh Program Studi Teknik Rekayasa Elektro-medis STIKes Al insyirah Pekanbaru, Jalan Parit indah No. 38
  • Yeni Pertiwi Program Studi Teknik Rekayasa Elektro-medis STIKes Al insyirah Pekanbaru, Jalan Parit indah No. 38
  • Romi Mulyadi Program Studi Teknik Rekayasa Elektro-medis STIKes Al insyirah Pekanbaru, Jalan Parit indah No. 38

DOI:

https://doi.org/10.35328/kesmas.v11i1.2178

Keywords:

blood pressure monitor, calibration, the uncertainty of measurement

Abstract

Testing and calibration of medical devices are one of the efforts to ensure the quality of medical devices. Tests and calibrations have been carried out on the Blood Pressure Monitor (BPM) in the emergency room at one hospital. In this article, it will be explained how the stages in testing and calibrating the BPM tool are to get the results of the appropriate/unfit for the tool. It aims to provide knowledge and awareness to health workers as device users and patients on the implementation and importance of testing and calibration of medical devices. The method used is the Work Method from the Ministry of Health of the Republic of Indonesia in 2018. The test parameters consist of a physical examination test and the function of the tool as well as a Blood Pressure performance test at the systole pressure measurement points of 60, 80, 100, 120, 150, 200 mmHg and the diastole pressure point 30, 50, 65, 80, 100, and 150 mm Hg. The results of the physical examination and function tests were 100% good. The results of the blood pressure performance test 94% passed the test with the measurement uncertainty value below the tolerance threshold (± 5 mmHg). The 150 mmHg diastolic pressure measurement point has a measurement uncertainty value that exceeds the tolerance threshold so that the pressure point does not pass the test. Thus, the BPM medical device is declared suitable for use.

 

References

BIPM, IEC, IFCC, ILAC, ISO, IUPAC, IUPAP, & OIML. (2020). GUM modeling: Guide to the expression of uncertainty in measurement - Part 6: Developing and using measurement models. JCGM GUM-6, 1–103. https://www.bipm.org/en/publications/guides

Kardianto, K., Kristanti, K. H., Tiswati, K. A., & Dwihapsari, Y. (2019). Analisis Nilai Ketidakpastian dan Faktor Kalibrasi pada Alat Ukur Radiasi di Balai Pengamanan Fasilitas Kesehatan Surabaya. Jurnal Fisika Dan Aplikasinya, 15(2), 56. https://doi.org/10.12962/j24604682.v15i2.4698

Kemenkes RI. (2015). Peraturan Menteri Kesehatan Republik Indonesia Nomor 54 Tahun 2015.

Kemenkes RI. (2016). Peraturan Menteri Kesehatan Republik Indonesia Nomor 43 Tahun 2016.

Laban, E. P., & Pinzon, R. T. (2017). SELUK BELUK HIPERTENSI: Peningkatan Kompetensi Klinis Untuk Pelayanan Kefarmasian. Berkala Ilmiah Kedokteran Duta Wacana, 2(3), 489. https://doi.org/10.21460/bikdw.v2i3.74

Saguni, A. (2018). Metode kerja ini digunakan untuk menentukan nilai batasan parameter pengukuran. 70, 1–355.

Downloads

Published

2022-07-30

How to Cite

ANALISIS HASIL PENGUJIAN DAN KALIBRASI PARAMETER BLOOD PRESSURE PADA ALAT KESEHATAN BLOOD PRESSURE MONITOR. (2022). Al-Tamimi Kesmas: Jurnal Ilmu Kesehatan Masyarakat (Journal of Public Health Sciences), 11(1), 45-50. https://doi.org/10.35328/kesmas.v11i1.2178