ANALYSIS INFUSION PUMP CALIBRATION RESULTS MERK TERUMO TYPE TE-112

Authors

  • Yeni Pertiwi Department of Medical Electronics Engineering Technology, Al Insyirah School of Health Science Pekanbaru, Indonesia
  • Nur Hadziqoh Department of Medical Electronics Engineering Technology, Al Insyirah School of Health Science Pekanbaru, Indonesia
  • Romi Mulyadi Department of Medical Electronics Engineering Technology, Al Insyirah School of Health Science Pekanbaru, Indonesia
  • Rino Ferdian Surakusumah Department of Medical Electronics Engineering Technology, Al Insyirah School of Health Science Pekanbaru, Indonesia
  • Tata Yunita Ovtaria Department of Medical Electronics Engineering Technology, Al Insyirah School of Health Science Pekanbaru, Indonesia

Abstract

The infusion pump functions to periodically control and regulate the administration of substances that are liquid nutrients or drugs at the right dose into the patient's body in a certain amount through the blood vessels with moderate precision. The use of an infusion pump for a long time will affect the accuracy and accuracy of the tool, which can result in compression and thus disrupt the flow rate regulation. For this reason, the Infusion Pump must be calibrated periodically. Infusion Pump calibration is carried out using a direct measurement method by testing the parameters of the physical condition and function of the tool, measuring electrical safety and its flow rate. Based on the results of the research and analysis of the calculation of the uncertainty of the measurement, the parameters of the physical condition and function of the tool as well as the electrical safety are in 100% good condition, while the flow rate parameter is in a condition of 66.67%. After reviewing the calibration results of the three parameters, the total value was 83.33%. So that the Infusion Pump tool is in a USAGE-worthy status.

References

Ali, M. I. (2019). Designing a Low-Cost and Portable Infusion Pump. 2019 4th International Conference on Emerging Trends in Engineering, Sciences and Technology (ICEEST), 1–4.https://doi.org/10.1109/ICEEST48626.2019.8981680

Ameliasari, L. (2019). Pengembangan Alat Penghitung Tetesan Infus Pada Terapi Intravena Berbasis Mikrokontroler Arduino Nano Untuk Menghitung Tetesan Infus [Bachelor, Universitas Muhammadiyah Purwokerto].

http://repository.ump.ac.id/9450/

Mansour, M. E. (2021). Design of Low Cost Smart Infusion Pump. 2020 International Conference on Computer, Control, Electrical, and Electronics Engineering (ICCCEEE), 1–5. https://doi.org/10.1109/ICCCEEE49695.2021.9429591

Maghfiroh, A., Havilda, N., & Das, S. (2022). Development of Infusion Device Analyzer Equipped with Occlusion Detection and a Real-Time Parameters Monitoring on Computer System | Jurnal Teknokes. http://teknokes.poltekkesdepkes-sby.ac.id/index.php/Teknokes/article/view/192

Faqih, F. N. (2021). INFUS PUMP [Diploma, Universitas Widya Husada Semarang]. https://eprints.uwhs.ac.id/434/

Hamzah, H., Amir, A., Pirade, Y. S., & Masarrang, M. (2021). Rancang Bangun Sistem Infus Cerdas. Foristek, 11(1), 8–19.https://doi.org/10.54757/fs.v11i1.138

Bagaskara, M., Dwiono, W., & Hayat, L. (2020). Rancang Bangun Pengatur Infus Berbasis Mikrokontroller. Jurnal Riset Rekayasa Elektro, 2(2), Article 2. https://doi.org/10.30595/jrre.v2i2.8320

Jannah, N. (2020). Rancang Bangun Infusion Device Analyzer Dengan Parameter Flowrate [Skripsi, Poltekkes Kemenkes Surabaya]. https://doi.org/10/12.%20BAB%203.pdf

Halimah, S. (2020). Perancangan Infusion Device Analyzer dengan Parameter Occlusion Berbasis Mikrokontroler [D3, Universitas Muhammadiyah Yogyakarta]. https://doi.org/10/Lampiran.pdf

Kresnoadi, E. (2018). Kelebihan Cairan (Fluid Overload) dan Hubungannya dengan Kejadian Acute Kidney Injury (Aki). Jurnal Kedokteran, 7(3), 23–23.

Mulyono, M., Wardaningsih, P. W., & Nugroho, A. S. (2019). Sistem Informasi Penjadwalan Pemeliharaan Dan Kalibrasi Alat Kesehatan. Prosiding Seminar Nasional Widya Husada, 0, Article 0. http://journal.uwhs.ac.id/index.php/psnwh/article/view/250

Ramadhani, R., Syaifudin, & Sumber. (2020). Analisis Keakurasian Sensor Tekanan Pada Parameter Occlusion Infusion Device Analyzer 2 Channel. Prosiding Seminar Nasional Kesehatan Poltekkes Kemenkes Surabaya 2020, 2(1), Article 1. http://semnas.poltekkesdepkes-sby.ac.id/index.php/2020/article/view/310

Puspitosari, L. A., Syaifudin, & Sumber. (2020). Analisis Keakurasian Sensor Photodiode Pada Parameter Flowrate Infusion Device Analyzer 2 Channel. Prosiding Seminar Nasional Kesehatan Poltekkes Kemenkes Surabaya 2020, 2(1), Article 1. http://semnas.poltekkesdepkes-sby.ac.id/index.php/2020/article/view/311

Pudji, A., Maghfiroh, A., & Thongpance, N. (2021). Design an Infusion Device Analyzer with Flow Rate Parameters using Photodiode Sensor | Indonesian Journal of Electronics, Electromedical Engineering, and Medical Informatics. http://ijeeemi.poltekkesdepkes-sby.ac.id/index.php/ijeeemi/article/view/100.

Downloads

Published

2023-04-14